    Conditioning the High Field Platform

    3.2 -Make sure the vacuum space is well pumped.
    -Lock up the platform and energize the HV power supply.
    -Check that the Current Potentiometer on the HV Supply is set to below 1 turn The current setting is 0.83 turns, corresponding to a current limit of about 0.25 mA (read via needle or via epics).
    -Start a strip tool of the HV voltage, current and vacuum (IG3).
    -Set the BIAS to 30 kV and condition.