    Focusing the camera

    The procedure for focusing the camera BNMR and BNQR is very similar. For example, the BNQR camera is mounted on the upper semicircular plate (shown in the picture below) and well focused on the sample position with these settings: (i) Focus setting 0.67m, (ii) Aperture wide open, and (iii) The F number is 1.4. When refocusing start with large F number=16 which gives you better depth of field, and may help you find the right focus settings. The camera lens can be accessed through a small window(Access port), shown in the picture below

    \begin{figure}\centering