Next: Sodium Handling at the Up: Conditioning the High Field Previous: Method A Contents Method B-Turn the current limiting pot to some low value.-Test the conditioning. Set the set point to some low value (5kV) and wait a moment for the plasma to dissipate (and the current to drop to microamps). Slowly increase the voltage to test if it will hold. -Repeat the procedure w/ the solenoid on and at field. -Repeat the procedure with the solenoid and EL3 on. -Tuning Alkali-Metal Ion Beams Through the Polarizer[1]. BNMR Web Site 2006-06-29 |
|||||||||||||