    Method B

    -Turn the current limiting pot to some low value.
    -Test the conditioning. Set the set point to some low value (5kV) and wait a moment for the plasma to dissipate (and the current to drop to microamps). Slowly increase the voltage to test if it will hold.
    -Repeat the procedure w/ the solenoid on and at field.
    -Repeat the procedure with the solenoid and EL3 on.
    -Tuning Alkali-Metal Ion Beams Through the Polarizer[1].