Calibration of the Low Field Spec. Magnet (ILE2A1:HH)
Calibration of the Low Field Spec. Magnet (ILE2A1:HH)
Conversion from Current to Field
Enter either the current of the field, then click
outside
the text box.
H=A+B*I
A=0.175+-0.046 G; B=2.2131+-0.0019
G/Amp
where H is
the field in
G and I is
the setpoint
current in Amps. |
Details of Calibration
The magnet is controlled by EPICS, where the setpoint is sent in Amps.
The read back from this magnet deviates from the setpoint linearly with
the value of the setpoint:
Fig. 1: The readback as a function of the setpoint of the
current in the magnet. The linear fit is
R=A+B*S, A=-0.036+-0.014
Amp, B=1.106+-4.4e-4
R is
readback and S
is setpoint. (Download
data)
The actual value of the
current that
runs in the magnet is closer to the setpoint that the read back, so
this is the value that we will use to calibrate the magnet, i.e. conver
current in Amps to field in Gauss. The calibration itself was performed
by looking at 8Li resonance in an annealed Au
foil as a
function of
different currents (year: 2002, RUNS: 45099-45103, 45110).
The resonance lines where found to be Gaussian!, to be more exact two
Gaussian lines. The center of the line was taken from the narrow line
of the two. The center of the narrow line as a function of the setpoint
current is presented in Fig. 2.
Fig. 2: The center of the narrow line as a function of the
setpoint current. The linear fit is
C=A+BI, A=0.11+-0.03 Amp; B=1.395+-0.0012
kHz/Amp
C is the
center in kHz and I
is the current in Amps. (Download data)
To get the value of the
field in G
corresponding to a setpoint in Amps, we devide the center by the
giromagnetic ratio of 8Li, γ=0.63018
kHz/G.
Fig. 3: The field in G as a function of the setpoint in
Amps.
The linear fit is:
H=A+B*I, A=0.175+-0.046
G; B=2.2131+-0.0019
G/Amp
H is the
field in G and I is
the setpoint current in Amps.
(Download
data)
To summarise the magnetic field as a function of the setpoint current is
H=A+B*I
A=0.175+-0.046 G; B=2.2131+-0.0019
G/Amp
where H is
the field in
G and I is
the setpoint
current in Amps. |
The applied magnetic field in bNQR from the Helmholtz coil is
perpendicular to the incident ion beam. Thus, as the ion beam enters
the field, it suffers a vertical deflection due to the Lorentz Force.
Actually the ions will make a a circular path with a certain radius of
curvature R, but R is much bigger than the size of the magnet, so the
deflection is rather small (as the field is not very large) but it is
significant for small samples. The attached plot shows a calibration of
the vertical beamspot deflection (measured using the scinitillator and
camera) as a function of field. The shift can be up or down depending
on the polarity of the magnet. To reverse the polarity one has to
reverse the leads physically on the power supply. Currently the
deflection is downwards (Oct.2006). In order to compensate for the
deflection, we can 1) move the cryostat or 2) have a different tune for
each field with the electrostatic correction benders compensating. We
are currently using method 1. The calibration plot allows this to be
done easily, using a reference point on the camera image. Note that
when we begin to decelerate on bNQR, this problem will be much more
serious, as R is proportional to (velocity/magnetic field).
Bending of Beamspot
(1)Zero Field (ZF) Beam-Spot
(2)150G Beam-Spot
|